Cod. EQCAP01
Cod. EQCON
Cod. TPRFR
Cod. TFRLANC
Cod. TFRC20
Cod. TFRMF38
Cod. TIND
Cod. TMPRF
Cod. TSTOPMFXX
(06/08/10/12/14/16)
Cod. TCREQ
For the correct insertion of the monophasic Equator Ø 3.8 implant, we recommend following the surgical protocol:
1. Use the lanceolate drill (TFRLANC) up to the marking (recommended n. revs = 800 rpm).
2. Use the 2 mm start-up drill (TFRC20) with the appropriate stop of the same height as the implant being inserted (recommended n. revs = 800 rpm).
3. Insert the parallelism indicator and check for any disparallelism.
4. Use the Ø 3.3 - Ø 3.8 drill (TFRMF38) with the appropriate stop (recommended n. revs = 200-400 rpm).
5. Insert the implant. Recommended n. of revs 15-25 rpm.