Monophasic Ø5.0

TSMM5006 for height of 6 mm
TSMM5008 for height of 8 mm
TSMM5010 for height of 10 mm
TSMM5012 for height of 12 mm
TSMM5014 for height of 14 mm
TSMM5016 for height of 16 mm

The height refers to the implantable part, to which the emerging part should be added, measuring 11.5 mm.

Drill extension

Cod. TPRFR

Lanceolate drill

Cod. TFRLANC

Cylindrical drill

Cod. TFRC20

Fresa per impianto monofasico Ø5,0

Cod. TFRMF50

Parallelism indicator

Cod. TIND

Depth gauge

Cod. TMPRF

Stop for drill for monophase implant

Cod. TSTOPMFXX
(06/08/10/12/14/16)

Ratchet for monophase implant

Cod. TCRICMF

Driver for monophase implant

Cod. TDRMF

Driver for monophase implant

Cod. TCRMF

Endoral welder

Surgical sequence of monophasic Ø 5.0 implant

For the correct insertion of the monophasic Ø 5.0 implant, we recommend following the surgical protocol:

1. Use the lanceolate drill (TFRLANC) up to the marking (recommended n. revs = 800 rpm).
2. Use the 2 mm start-up drill (TFRC20) with the appropriate stop of the same height as the implant being inserted (recommended n. revs = 800 rpm).
3. Insert the parallelism indicator and check for any disparallelism.
4. Use the Ø 5.0 drill (TFRMF45) with the appropriate stop (recommended n. revs = 200-400 rpm).
5. Hold the implant using the cap of the vial where the implant itself is anchored and insert it into the surgical alveolus taking care to avoid any contact with non-sterile surfaces. Partially screw the implant using the cap then remove this cap and insert the sterile TCHMF wrench to complete screwing.
6. Insert the implant. Recommended n. of revs 15-25 rpm.